CVE-2018-18563

9.6 · CRITICAL
Published Nov 20, 2018 roche CWE-434 EPSS 0.97% (59th pctl)

Overview

CVE-2018-18563 is a critical-severity vulnerability affecting roche accu-chek_inform_ii_firmware. It was published on November 20, 2018 and has a CVSS 3.0 base score of 9.6 (CRITICAL).

This vulnerability has a CVSS 3.0 base score of 9.6, rated CRITICAL. It requires local or adjacent network access to exploit. No authentication or special privileges are required for exploitation.

Technical Description

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.

Remediation

Check the references section for vendor advisories and patches from roche. Update accu-chek_inform_ii_firmware to the latest patched version. If immediate patching is not possible, review the CVSS vector to understand the attack surface and apply compensating controls such as network segmentation or access restrictions.

Affected Products

Vendor Product Versions Status
roche accu-chek_inform_ii_firmware >= 0, < 03.06.00 Affected
roche cobas_h_232_firmware >= 0, < 03.01.03 Affected
roche coaguchek_pro_ii_firmware >= 0, < 04.03.00 Affected
roche coaguchek_xs_plus_firmware >= 0, < 03.01.06 Affected
roche coaguchek_xs_pro_firmware >= 0, < 03.01.06 Affected

Frequently Asked Questions

What is CVE-2018-18563?

CVE-2018-18563 is a critical-severity vulnerability affecting roche accu-chek_inform_ii_firmware. It was published on November 20, 2018 and has a CVSS 3.0 base score of 9.6 (CRITICAL).

How severe is CVE-2018-18563?

This vulnerability has a CVSS 3.0 base score of 9.6, rated CRITICAL. It requires local or adjacent network access to exploit. No authentication or special privileges are required for exploitation.

How do I fix or remediate CVE-2018-18563?

Check the references section for vendor advisories and patches from roche. Update accu-chek_inform_ii_firmware to the latest patched version. If immediate patching is not possible, review the CVSS vector to understand the attack surface and apply compensating controls such as network segmentation or access restrictions.

How can CyberStrike help with CVE-2018-18563?

CyberStrike's AI-powered security agents can automatically detect CVE-2018-18563 across your infrastructure using autonomous pentesting, DAST scanning, and HackBrowser. The platform continuously monitors for known vulnerabilities and provides actionable remediation guidance prioritized by real-world exploitability.